Language:

MS-S150 Series Syringe Pump

MS-S150 Series Syringe Pump

Function and Features
1.Awarded with ISO13458:2003(TUV Germany)and CE Certification
2.All standard syringes of 10ml,20ml,50ml are compatible

3.The Machine diagnoses the syringe size automatically,and with a simple syringe specification setting. 
4.The syringe pump can guarantee an accurate and safe injection. 
5.Unique function of bedside injection supervision which makes patients and their relatives as well as nurses more assured. 
6.Excellent injection function and operation. Even small-volume injection is accurate. 
7.Double CPU system ensures safe and reliable injection process. 
8.Driven by precision step motor, no pulsating waves shall occur during even low-rate injection. 
9.Accurate and sensitive pressure sensor detects occlusion pressure accurately. 
10.Alarms are given in case of malfunction or operation error and the machine stops operation automatically. 

Technical Parameters


Applicable Syringes

All brands of stand syringes:

 

Weight

<3.0kg

MS-150A

10ml,20ml,50ml

Dimension

3801×901×80mm(lengthw×idthh×eight)

MS-150B

1ml,5ml,10ml

Power supply

 

AC100V~240V;0/60HZ

 Accuracy

±3%

Fuse

T2AL 250V

Injection rate

10ml:0.1~100ml/h

20ml:0.1~200ml/h

50ml:0.1~500ml/h

Maximum Power consumption

Battery

25VA, run for 4h after being charged.

Rechargeable lithium Polymer battery,7.4V,1600mAh

1ml:0.1~10ml/h

5ml:0.1~50ml/h

10ml:0.1~100ml/h

Battery recharge

When the pump is connected to the AC power supply, turn on the power and the battery will automatically recharge (about 8-14 hours to recharge fully.)

Accumulated  volume

0.1ml~999.9ml

Electrical safety

Comply with the requirement of IEC60601-1;1996

 

Increment

 

0.1ml/h

Electromagnetic compatibility

Comply with the requirement of IEC60601-2;2006

Bolus rate

 

 

 

 

MS-150A

The fastest rate of the size of syringe in use

Shell material

ABS plastic

Classification

Class II, Type BF

MS-150B

No bolus

Operating conditions

Environment temperature 5~45°C.air pressure 86kPa~106kPa, relative humidity≤80%

Displayed information

Rate, accumulated injection volume, Syringe size, battery capacity, bed No., AC power indicator, battery charge indicator

Storage conditions

Environment temperature -40°C~55°C, air pressure 50kPa ~106kPa, relative humidity ≤95%

Alarm functions

 

Over, Near, Occlusion, low battery, syringe disengaged, No AC power supply, Serious low battery

Applied standards

IEC60601-1: 2005

IEC60601-2-24: 1998

IEC60601-1-2: 2006

ISO780-1997

ISO15223: 2000

ISO7886-1: 1993

ISO / IEC: Guide No. 37<onsumer Goods Usage Instructions>