MS-W500S Infant Radiant Warmer

MS-W500S Infant Radiant Warmer
Brief introduction:

* Microprocessor based servo controlled temperature system.
* The microprocessor-controlled unit allows for three settings: pre-warm, manual or automatic.
* Alarm functions for safety includes:
Power failure;
Temperature deviation;
Temperature sensor failure;

Over temperature.
* APGAR timer to record therapy time
* Convenient X-ray cassette under infant bed
* Heater head can be adjusted +90º horizontally
* Tilt angle of infant bed is adjustable
* Observe light angle is adjustable
* Transparent protector folded
* Adopts anti-explosion micro crystal quartz infrared radiation tube for its radiation source,the heating is quick and life-span is long
* Alarms are audible and visual
* A second thermal cut-out function for more safety
* Skin temperature sensor failure protect function avoid over temperature
* Temperature deviation can be amended in front panel directly


Standard configuration:
Main body (including the radiant source, control system, infant bed, bracket), X-ray cassette , Observe light , I.V.pole, skin temperature sensor, tray, mattress, transparent protector, castors.
Optional parts:
Drawer;
RS-232(485) connector;


Technical data:
1.Power supply : 110/220V 60/50Hz
2.Power input: ≤1200VA
3.Operating condition:
An ambient Temperature:18°C -30°C
 An ambient Relative Humidity: 30%--75%RH
 Atmospheric Pressure: 700~1060 hPa
 An ambient Air Velocity < 0.3m/s
4. Temperature control: 32°C~38°C
5.Temperature control accuracy: ≤0.5°C
6.Mattress temperature uniformity : < 2°C
7.Skin temperature sensor accuracy: + 0.3°C
8.Warm-up time: ≤30min
9. Infant bed tilt angle:+ 10°
10.APGAR timer range: ≤24h
11.Heater head horizontal: ±90°
12.Mattress size: 67cm × 54cm
13.Distance from heater to mattress:80cm
Transport and storage:
Environment Temperature:-40°C ~ +55°C
Environment Relative Humidity: ≤95%
Atmospheric Pressure: 500~ 1060hPa


Package:
Each unit is packed one case
Size:120cm× 70cm × 96cm
Gross weight: ≤90kg